S100(300)-2B+mYZ15

Flow range: 0.0048-440 mL/min; single-channel and dual-channel pump heads are optional

    S100(300)-2B+mYZ15

    • Product Description
    • Commodity name: S100(300)-2B+mYZ15

    Flow range: 0.0048-440 mL/min; single-channel and dual-channel pump heads are optional

    Product Introduction

    Flow range: 0.0048-440mL/min

    Single and dual channel pump heads are optional

    3-roller and 6-roller pump heads are optional. 3-roller pump heads have a larger flow rate, while 6-roller pump heads have lower pulsation.

    Maximum speed of 100rpm and 300rpm are optional

    Stepper motor drive

    15-digit BCD DIP switch speed control, or external control signal speed control

    On/off control by switch quantity, forward/reverse control by switch quantity

    Flow accuracy can reach ±1%

    Panel or base mounting

    24V DC power supply, optional power adapter for 220V AC power supply

    Tube Installation

    Turn the duckbill wrench counterclockwise by 180°, and lift the pressure block

    Insert the tube naturally between the roller and the upper pressure block

    Turn the duckbill wrench clockwise by 180°, press the pressure block into place, and complete the tube installation.

    Technical Parameters

    Product Model S100(300)-2B+mYZ15
    Applicable Pump Head mYZ15-13mYZ15-16mYZ15-23mYZ15-26
    Start/Stop Control Can control start and stop
    Direction Control Reversible
    Internal Speed Control BCD DIP switch adjustment
    External Speed Control 0-10kHz pulse control or 0-5V, 0-10V, 4-20mA analog signal control, choose one. Standard configuration is 0-10KHz.
    S100-2B Internal Speed Range 1-100rpm, 15-speed adjustment, standard configuration is 5,10,15,20,25,30,35,40,45,50,60,70,80,90,100rpm, customizable
    S100-2B External Speed Range 0-100rpm, speed resolution 0.1rpm, full range linear correspondence
    S300-2B Internal Speed Range 1-300rpm, 15-speed adjustment, standard configuration is 20 40 60 80 120...280 300rpm, customizable
    S300-2B External Speed Range 0-300rpm, speed resolution 0.1rpm, full range linear correspondence
    Communication Control RS485 communication control, using Modbus communication protocol. Communication control and other external control cannot coexist, choose one.
    Speed Error ≤±0.5%
    Start/Stop Control Start/Stop Control:Normally open switch quantity control contact, peristaltic pump runs; contact closed (port 5 and port 7 short-circuited), peristaltic pump stops.
    Direction Control Direction Control:Normally open switch quantity control contact, peristaltic pump runs clockwise; normally closed contact (port 6 and port 7 short-circuited), peristaltic pump runs counterclockwise.
    Applicable Power Supply 11.4-25.2V DC; or 100-240V AC (optional power adapter) 50/60Hz
    Operating Environment 0-40℃, Relative Humidity: <80%

    Selection Guide

    Peristaltic Pump Model Speed (rpm) Number of Rollers Number of Channels Applicable Tubing Flow Range (ml/min)
    S100-2B+mYZ15-13 10031

    13#14#19#

    16#25#

    0.006-138
    S100-2B+mYZ15-16 61 0.0048-90
    S100-2B+mYZ15-23 32

    13#14#

    19#16#

    0.006-59.4
    S100-2B+mYZ15-26 62 0.0048-46.5
    S300-2B+mYZ15-13 30031

    13#14#19#

    16#25#

    0.006-440
    S300-2B+mYZ15-16 61 0.0048-257
    S300-2B+mYZ15-23 32

    13#14#

    19#16#

    0.006-205.9
    S300-2B+mYZ15-26 62 0.0048-144.2

    Dimensions

     S100-1B+MYZ15 Dimensions

Get a free product quote

Fill in your phone and E-mail information, and we will contact you within one working day to solve your problems as soon as possible.