114DV, 116DV

Flow rate range: 0.005-560mL/min, applicable speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

Category:

    114DV, 116DV

    • Product Description
    • Commodity name: 114DV, 116DV

    Flow rate range: 0.005-560mL/min, applicable speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

    Product Introduction

    Flow Range: 0.005-560mL/min

    Applicable Speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

    The 114DV (4 rollers) pump head has a larger flow rate, while the 116DV (6 rollers) pump head has lower pulsation.

    Flip-top design, compact size, convenient and reliable operation
    Elastic compression tubing, long tubing life

    Selection Guide

    Pump Head Model External Material Roller Material Number of Pump Head Rollers Applicable Tubing Applicable Speed (RPM) Reference Flow Rate (mL/min) Weight (KG)
    114DV IXEF PET+TX 4 13#,14#,19#,16#,25# ≤6005600.2
    116DV6≤4002260.2

    Tubing Installation

     114DV Tubing Installation

    Reference Flow Rate

    114DV Reference Flow Rate

    Speed Average Flow Rate ml/min Speed Average Flow Rate ml/min
    rpm 13# 14# 19# 16# 25# rpm 13# 14# 19# 16# 25#
    100.471.422.935.089.331506.9921.343.9576.2139.95
    200.932.845.8610.1618.662009.3228.458.6101.6186.6
    301.404.268.7915.2427.9925011.6535.573.25127233.25
    401.865.6811.7220.3237.3230013.9842.687.9152.4279.9
    502.337.1014.6525.4046.6535016.3149.7102.55177.8326.55
    602.808.5217.5830.4855.9840018.6456.8117.2203.2373.2
    703.269.9420.5135.5665.3145020.9763.9131.85228.6419.85
    803.7311.3623.4440.6474.6450023.3071146.5254466.5
    904.1912.7826.3745.7283.9755025.6378.1161.15279.4513.15
    1004.6614.229.350.893.360028.0085176305560

    Dimensions

Get a free product quote

Fill in your phone and E-mail information, and we will contact you within one working day to solve your problems as soon as possible.