114DV, 116DV

Flow rate range: 0.005-560mL/min, applicable speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

Category:

    114DV, 116DV

    • Product Description
    • Commodity name: 114DV, 116DV

    Flow rate range: 0.005-560mL/min, applicable speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

    Product Introduction

    ◆ Flow Range: 0.005-560mL/min

    ◆ Applicable Speed: ≤400rpm (continuous operation), ≤600rpm (intermittent operation)

    ◆ The 114DV (4 rollers) pump head has a larger flow rate, while the 116DV (6 rollers) pump head has lower pulsation.

    ◆ Flip-top design, compact size, convenient and reliable operation
    ◆ Elastic compression tubing, long tubing life

    Selection Guide

    Pump Head Model External Material Roller Material Number of Pump Head Rollers Applicable Tubing Applicable Speed (RPM) Reference Flow Rate (mL/min) Weight (KG)
    114DV IXEF PET+TX 4 13#,14#,19#,16#,25# ≤6005600.2
    116DV6≤4002260.2

    Tubing Installation

     114DV Tubing Installation

    Reference Flow Rate

    ◆ 114DV Reference Flow Rate

    Speed Average Flow Rate ml/min Speed Average Flow Rate ml/min
    rpm 13# 14# 19# 16# 25# rpm 13# 14# 19# 16# 25#
    100.471.422.935.089.331506.9921.343.9576.2139.95
    200.932.845.8610.1618.662009.3228.458.6101.6186.6
    301.404.268.7915.2427.9925011.6535.573.25127233.25
    401.865.6811.7220.3237.3230013.9842.687.9152.4279.9
    502.337.1014.6525.4046.6535016.3149.7102.55177.8326.55
    602.808.5217.5830.4855.9840018.6456.8117.2203.2373.2
    703.269.9420.5135.5665.3145020.9763.9131.85228.6419.85
    803.7311.3623.4440.6474.6450023.3071146.5254466.5
    904.1912.7826.3745.7283.9755025.6378.1161.15279.4513.15
    1004.6614.229.350.893.360028.0085176305560

    Dimensions

Get a free product quote

Fill in your phone and E-mail information, and we will contact you within one working day to solve your problems as soon as possible.