LSP01-2A Micropump
Category:
Tags:
- Product Description
LSP01-2A Micropump
-
- Commodity name: LSP01-2A Micropump
Flow rate range: 0.0002 ul/min-8.6699 ml/min. Single-channel infusion pump; micro-injection pump suitable for high-precision, low-flow liquid transfer.
Product Introduction
◆ The LSP01-2A micro-syringe pump is a single-channel perfusion syringe pump, mainly used in biological laboratories, with an integrated structure.
◆ Able to accommodate syringes (injectors) of different specifications, the LSP01-2A micro-syringe pump is suitable for high-precision, low-flow liquid transfer.
◆ Flow rate range: 0.0002ul/min-8.6699ml/min.
◆ LSP01-2A syringe inner diameter input function: You can select the syringe from the list or directly input the syringe inner diameter.
◆ This micro-syringe pump has a user-friendly interface: large-screen LCD display, digital rotary knob and membrane buttons are used together, making operation simple and quick.
◆ Working mode: Single-push working mode.
◆ Power-off memory function: Saves the set parameters, and no need to reset after power-on.
◆ Jam protection function: When the injection pump's drive mechanism is blocked during operation, the injection pump will stop the drive mechanism's work and issue a buzzer alarm.
◆ Using a 485 communication bus, it can be connected to a host computer for control through background software.
◆ External control interface: Has input/output control functions.
◆ Liquid volume calibration function: A more accurate flow rate can be obtained through the calibration program.
◆ Syringe protection function: By adjusting the position of the limit block, the syringe can be prevented from being damaged.
Typical Applications
◆ Biological laboratory
◆ Electrophysiology, chemistry, pharmaceuticals, etc.
◆ Micro liquid perfusion
◆ Electrospinning
Control Panel
Technical Parameters
Product Model LSP01-2A Micro-syringe pump working mode Perfusion Flow Rate Range 0.0002ul/min-8.6699ml/min Number of syringes that can be installed 1 Stroke Distance 140mm Stroke Resolution 0.03125μm Linear Velocity Range 1μm/min-13mm/min (Flow rate = Linear velocity × Syringe inner cross-sectional area) Linear Velocity Adjustment Resolution 1μm/min Stroke Control Accuracy Error ≤ ±0.5% (when stroke ≥ 30% of maximum stroke) Rated Linear Thrust >90N Syringe Selection Built-in syringes from major manufacturers and models for selection Syringe Customization You can directly input the inner diameter of the syringe Flow Rate Calibration Obtain a more accurate liquid volume through the calibration program Running Parameter Setting Dispensing volume, injection time, etc. Display Parameter Selection Volume, flow rate, or linear velocity Power-off Memory You can choose whether to continue working according to the status before power-off after power-on Status Signal Output 2-channel OC gate signal output, used to indicate start/stop and direction status Control Signal Input 2-channel start/stop control input terminals, 1-channel falling edge trigger signal controls start/stop; 1-channel TTL level signal controls start/stop Communication Interface RS485 Dimensions 280×210×140(mm) Weight 3.6kg Power Supply AC 90V-260V/15W Working Environment Relative Humidity <80% Working Environment Temperature 0℃-40℃ Selection Guide
Syringe Pump Model
Applicable Syringe (or Injector) Specifications
Applicable Syringe Inner Diameter (mm)
Reference Flow Rate Range (μl/min-ml/min)
LSP01-2A
10μl
0.50
0.0002-0.0026
25μl
0.80
0.0005-0.0065
50μl
1.10
0.0095-0.0124
100μl
1.60
0.002-0.026
250μl
2.30
0.0042-0.054
500μl
3.25
0.0083-0.1078
1ml
4.72
0.0175-0.2275
2ml
9.00
0.0636-0.827
5ml
13.10
0.1348-1.7522
10ml
16.60
0.2164-2.8135
20ml
19.00
0.2835-3.6859
30ml
23.00
0.4155-5.4012
60ml
29.14 0.6669-8.6699
Dimensions
Related products
Get a free product quote
Fill in your phone and E-mail information, and we will contact you within one working day to solve your problems as soon as possible.